Computer-aided evaluation of the anatomical accuracy of hybrid SPECT/spiral-CT imaging of lesions localized in the neck and upper abdomen

Riferimento: 
Nucl Med Commun. 2012 Nov;33(11):1153-9.
Autori: 
Bennewitz C, Kuwert T, Han J, Ritt P, Hahn D, Thimister W, Hornegger J, Uder M, Schmidt D.
Fonte: 
Nucl Med Commun. 2012 Nov;33(11):1153-9.
Anno: 
2012
Azione: 
Precisione anatomica della tomografia computerizzata con 111In-octreotide nella parte superiore dell'addome e del collo.
Target: 
111In-octreotide/addome superiore e collo.

ABSTRACT
OBJECTIVE:
The purpose of this study was to investigate the anatomical accuracy of hardware-based single-photon emission computed tomography/computed tomography (SPECT/CT) registration in the upper abdomen and neck.
METHODS:
The database consisted of 90 patients referred for SPECT/CT for diagnostic workup of either thyroid/parathyroid disease (n=46) or abdominal neuroendocrine tumours (n=44). In the first group, 99mTc-MIBI was used as the tracer and in the second 123I-metaiodobenzylguanidine (n=13), 111In-octreotide (n=28) or 99mTc-octreotide (n=3). For predefined structures represented by both modalities, the distances between the centres of gravity of their CT and SPECT representation were determined in a semiautomated manner. In cervical data sets, this analysis was performed for the submandibular salivary glands (n=92) and in abdominal data sets for 69 neoplastic foci.
RESULTS:
The mean distances were 5.7 ± 2.0 mm (range: 1.84-9.67 mm) in the neck and 6.8 ± 3.3 mm (range: 1.4-19.7 mm) in the abdomen. In 42 out of 92 of the cervical and 40 out of 69 of the abdominal data sets at least one of the X-direction-determined, Y-direction-determined, and Z-direction-determined distances was greater than the SPECT pixel width of 4.6 mm.
CONCLUSION:
The anatomical accuracy of hardware-based SPECT/CT fusion depends also on the region of the body studied. For example, in the neck and upper abdomen the accuracy is lower than in the lower lumbar spine. In clinical routine, SPECT/CT data sets acquired for the neck and upper abdomen should be regularly checked and corrected for SPECT/CT misalignment. This is, in particular, important when CT-based corrections of SPECT involving pixelwise data integration such as for attenuation correction are made.